ANSI/ASHRAE 110, Method of Testing Performance of Laboratory Fume Hoods, is the reference method most U.S. laboratories, universities and pharmaceutical sites use to evaluate whether a fume hood contains what is generated inside it. It is a method rather than a pass/fail standard: it tells a tester how to measure and leaves the acceptance criteria to the owner, the design engineer or the applicable code.

What the test actually involves

A complete ASHRAE 110 evaluation has four parts, and each answers a different question about the hood.

  • Inspection and setup. The hood, its exhaust, its sash and the surrounding room are documented before anything is measured, including sash position, room conditions, supply air and nearby disturbances.
  • Airflow visualization. A smoke source shows how air moves into, across and out of the hood, and where it escapes. It is the fastest way to see a cross-draft, a dead zone behind equipment or a leaking sash gasket.
  • Face velocity. Averaged and individual readings are taken across the sash opening. Velocity tells you how much air is entering, but on its own it does not tell you how well the hood contains.
  • Tracer gas containment. A controlled release of tracer gas inside the hood is measured at the breathing zone of a mannequin standing at the face and in the room. This is the part that measures containment rather than airflow.

Why face velocity is not containment

Higher velocity is not automatically better. Above roughly 100 feet per minute, eddy currents form around the tester's body and can pull contaminants back toward the breathing zone, and excessive turbulence at the sash opening does the same. Supply diffusers aimed at the hood, foot traffic, a nearby door or equipment stacked inside the work area can defeat a hood that reads perfectly on a velocity meter. The tracer gas test exists because those effects do not appear in a velocity number.

What AM, AI and AU mean

ASHRAE 110 results are often reported as a rating rather than a single number. AM means as manufactured, AI means as installed and AU means as used, with the user's equipment in place. The rating summarizes both the containment level and how stable containment remained when the tester walked past or moved in front of the hood. A hood that contains well when undisturbed but loses control with movement is a different problem from one that never contains at all, and the report should say which condition you have.

Who decides pass or fail

Because the standard is a method, the acceptance criteria come from somewhere else: the owner's EHS program, the design specification, a university standard, NFPA 45 or a manufacturer's published claim. A report should state the criteria it was measured against, so a later reviewer can see whether the hood met the owner's requirement or only the tester's expectation.

When to test

  • On installation, before the hood is placed into service
  • After relocation, exhaust or control work, or a change to the sash or baffle configuration
  • At the interval set by the facility program, commonly annually and more often for high-hazard work
  • When users report odors, alarms or symptoms suggesting the hood is not containing

Where the hood fits in the laboratory

A fume hood is one component of a ventilation system. Room pressure relative to the corridor, supply air delivered near the hood face, exhaust capacity shared with other hoods and the response of the air valve all affect the result. That is why testing a hood in isolation, before the room is balanced, produces a result that cannot be defended later. Testing once the room is stable, and documenting those conditions, is what makes the number meaningful.

Fume hood performance testing from Micron

Micron performs fume hood performance testing, including ASHRAE 110 methods within the agreed scope, alongside room airflow and pressure verification when the project calls for it. If a hood reads fine but does not seem to protect, or you have a certification report you could not defend to an auditor, start with the scope: which tests, at what sash position, against which criteria.